New Estimators and Guidelines for Better Use of Fetal Heart Rate Estimators with Doppler Ultrasound DevicesReportar como inadecuado




New Estimators and Guidelines for Better Use of Fetal Heart Rate Estimators with Doppler Ultrasound Devices - Descarga este documento en PDF. Documentación en PDF para descargar gratis. Disponible también para leer online.

Computational and Mathematical Methods in Medicine - Volume 2014 2014, Article ID 784862, 10 pages -

Research Article

Signal & Imaging Group, University François Rabelais of Tours, PRES Loire Valley University, UMR INSERM U930, 7 Avenue Marcel Dassault, 37200 Tours Cedex, France

Inserm, U930, 10 Boulevard Tonnellé, BP 3223, 37032 Tours Cedex, France

IRIT UMR 5505, Université Paul Sabatier Toulouse 3, 118 Route de Narbonne, 31062 Toulouse Cedex 9, France

Received 28 May 2013; Revised 25 October 2013; Accepted 4 November 2013; Published 29 January 2014

Academic Editor: Brynjar Karlsson

Copyright © 2014 Iulian Voicu et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Characterizing fetal wellbeing with a Doppler ultrasound device requires computation of a score based on fetal parameters. In order to analyze the parameters derived from the fetal heart rate correctly, an accuracy of 0.25 beats per minute is needed. Simultaneously with the lowest false negative rate and the highest sensitivity, we investigated whether various Doppler techniques ensure this accuracy. We found that the accuracy was ensured if directional Doppler signals and autocorrelation estimation were used. Our best estimator provided sensitivity of 95.5%, corresponding to an improvement of 14% compared to the standard estimator.





Autor: Iulian Voicu, Sébastien Ménigot, Denis Kouamé, and Jean-Marc Girault

Fuente: https://www.hindawi.com/