Abstract

Toxic anterior segment syndrome TASS developed in four cases after uneventful implantation of a foldable iris-fixated phakic intraocular lens pIOL. Two cases occurred sequentially in one patient. The TASS subsided without complications in all cases after intensive topical steroid treatment. A multitude of possible causes is considered for the occurrence of these TASS cases. From the sterilization and cleaning of surgical instruments to the possibility of endotoxines in ophthalmic viscosurgical devices OVD. These rare cases should alert the surgeon to the possibility of TASS after pIOL implantation.