Accuracy of optical biometry combined with Placido disc corneal topography for intraocular lens power calculationReport as inadecuate




Accuracy of optical biometry combined with Placido disc corneal topography for intraocular lens power calculation - Download this document for free, or read online. Document in PDF available to download.

Purpose

To investigate the accuracy of a new optical biometer for intraocular lens IOL power calculation in eyes undergoing cataract surgery.

Methods

Consecutive eyes of patients undergoing cataract surgery with the same IOL model were enrolled in a prospective cohort study.Axial length AL and corneal power were measured with an optical biometer based on optical low-coherence interferometry and Placido-disc corneal topography. IOL power was calculated with the Hoffer Q, Holladay 1 and SRK-T formulas. For each formula the lens constant was optimized in retrospect in order to achieve a mean prediction error PE of zero difference between the predicted and the postoperative refraction. Median absolute error MedAE and percentage of eyes with PE ±0.50 D were calculated.

Results

Seventy-four eyes of 74 cataract patients were enrolled. The MedAE was 0.25 D with all formulas. A PE within ±0.50 D was obtained in 89.04% of cases with the Hoffer Q and SRK-T formulas, and in 87.67% of cases with the Holladay 1 formula.

Conclusions

The optical biometer investigated in the present study provides accurate measurements for IOL power calculation.



Author: Giacomo Savini , Kenneth J. Hoffer, Piero Barboni, Nicole Balducci, Domenico Schiano-Lomoriello, Pietro Ducoli