Feasibility of rapid quantitation of stratum corneum lipid content by Fourier transform infrared spectrometryReportar como inadecuado




Feasibility of rapid quantitation of stratum corneum lipid content by Fourier transform infrared spectrometry - Descarga este documento en PDF. Documentación en PDF para descargar gratis. Disponible también para leer online.

Spectroscopy - Volume 18 2004, Issue 3, Pages 423-431



Institute of Clinical Pharmacy, College of Medicine, National Cheng Kung University, Tainan 701, Taiwan

Department of Pharmacy, Chia-Nan University of Pharmacy and Science, Tainan Hsien 717, Taiwan

Department of Dermatology, College of Medicine, National Cheng Kung University, 1 University Rd., Tainan 701, Taiwan

Department of Chemical Engineering, College of Engineering, National Cheng Kung University, 1 University Rd., Tainan 701, Taiwan

Pfizer Limited, Tamsui, Taiwan



Copyright © 2004 Hindawi Publishing Corporation. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

The permeability barrier of skin resides in the stratum corneum, and its properties are mediated by a series of lipid multilayers, enriched in ceramides, cholesterol, and free fatty acids, segregated within the stratum corneum SC interstices. SC lipid content is usually determined by gravimetric methods in conjunction with high performance thin layer chromatography, but these methods are time‒consuming and involve hazardous solvents. The objective of the present study was to develop a method of measuring SC lipid content by Fourier transform infrared spectrometry FTIR that is fast and requires no solvents. The IR spectra of isolated porcine SC sheets were recorded using a FTIR spectrometer. SC lipid content was determined by gravimetric methods using chloroform–methanol extraction. The peak area of both the CH2 symmetric 2850 cm

and asymmetric 2920 cm

stretching bands in the IR spectra of progressively solvent‒extracted porcine SC sheets decreased with increasing amount of SC lipids removed. When spectral analysis was performed by curve‒fitting using GRAMS-32 software between 3000 to 2800 cm

, peak area ratios of CH2 to CH3 asymmetric stretching bands in the IR spectra of 46 isolated porcine SC samples were correlated to SC lipid content R

=0.90, with the standard error of measurement of 1.91%. The study demonstrated the feasibility of using FTIR technique to rapidly and accurately measure SC lipid content.





Autor: Jui-Chen Tsai, Yu-Li Lo, Ching-Yu Lin, Hamm-Ming Sheu, and Jui-Che Lin

Fuente: https://www.hindawi.com/