* Corresponding author 1 LPL - Laboratoire Parole et Langage

Abstract : To measure oral and nasal airflow during speech production, several conditions must be gathered. In this aim, we designed and built a pneumotachograph with particular care to optimise its response time, linearity and acoustical response. This flow meter is based on the grid flow meter principle with a small dead volume and specific linearization for the inhaled and exhaled airflow. A soft silicone rubber mask, pressed against the speaker-s face prevents air leakage, without hindering articulatory movements. The acoustical distortions of the speech sound through the device are remedied by an adapted signal processing from its transfer function.

Keywords : speech and voice assessment airflow pneumotachograph aerodynamic





Autor: Alain Ghio - Bernard Teston -