Low-Dose High-Pitch CT Angiography of the Supraaortic Arteries Using Sinogram-Affirmed Iterative ReconstructionReportar como inadecuado




Low-Dose High-Pitch CT Angiography of the Supraaortic Arteries Using Sinogram-Affirmed Iterative Reconstruction - Descarga este documento en PDF. Documentación en PDF para descargar gratis. Disponible también para leer online.

Objective

To prospectively evaluate image quality and radiation dose using a low-dose computed tomography angiography protocol and iterative image reconstruction for high-pitch dual-source CT-angiography DSCTA of the supraaortic arteries.

Material and Methods

DSCTA was performed in 42 patients, using either 120 kVp tube voltage, 120 mAS tube current, 2.4 pitch and filtered back projection, or 100 kVp tube voltage, 100 mAs tube current, 3.2 pitch, and sinogram affirmed iterative reconstruction. Measurements of vessel attenuation, of the contrast-to-noise ratio CNR and the signal-to-noise ratio SNR were performed to objectively evaluate image quality. Two readers evaluated subjective image quality and image noise, using a four-point scale. Effective dose was used to compare the differences in radiation dose.

Results

Low-dose protocol application showed significantly higher vessel opacification p = 0.013, and non-significantly higher CNR and SNR values. There was no difference in the subjective image quality and image noise reading between the protocols. Effective dose was significantly lower using the low-dose protocol 1.29±0.21 mSv vs. 2.92±0.72 mSv; p<0.001.

Conclusion

The combined use of reduced tube voltage, reduced tube current, and iterative reconstruction reduces radiation dose by 55.4% in high-pitch DSCTA of the supraaortic arteries without impairment of image quality.



Autor: Dietrich Beitzke , Richard Nolz, Sylvia Unterhumer, Christina Plank, Michael Weber, Rüdiger Schernthaner, Veronika Schöpf, Flor