A GATE Simulation Study of the Siemens Biograph DUO PET-CT SystemReportar como inadecuado




A GATE Simulation Study of the Siemens Biograph DUO PET-CT System - Descarga este documento en PDF. Documentación en PDF para descargar gratis. Disponible también para leer online.

This is a GATE-simulation study of the Siemens BiographDUO PET-CT system. It reports effects of changes in the thickness of theemployed Lutetium OxyorthosilicateLSO detectors. The PET-CT, a human bodyphantom and a cylindrical F-18 FDG source were simulated. Validationmeasurements were conducted. The results indicate that LSO thickness increase degradesspatial resolution, improves relative energy resolution from 9.0% to 11.3% andincreases signal-to-noise-ratio from 0.81 to 1.17. Thicker LSO crystals presentgreater axial sensitivity so as the detection efficiency of PET would besignificantly enhanced.

KEYWORDS

Monte-Carlo; GATE; PET; Siemens Biograph DUO PET-CT

Cite this paper

D. Nikolopoulos, S. Kottou, N. Chatzisavvas, X. Argyriou, E. Vlamakis, P. Yannakopoulos and A. Louizi -A GATE Simulation Study of the Siemens Biograph DUO PET-CT System,- Open Journal of Radiology, Vol. 3 No. 2, 2013, pp. 56-65. doi: 10.4236-ojrad.2013.32009.