Multiple Silicon Nanowires with Enzymatic Modification for Measuring Glucose ConcentrationReportar como inadecuado




Multiple Silicon Nanowires with Enzymatic Modification for Measuring Glucose Concentration - Descarga este documento en PDF. Documentación en PDF para descargar gratis. Disponible también para leer online.

1

Department of Photonics Engineering, Yuan Ze University, 135, Yuan-Tung Road, Chung-Li 32003, Taiwan

2

Department of Mechanical and Electromechanical Engineering, Tamkang University, New Taipei 25137, Taiwan



These authors contributed equally to this work.





*

Author to whom correspondence should be addressed.



Academic Editor: Joost Lötters

Abstract This study fabricated a multiple poly-Si nanowires sensor through a top-down method and immobilized glucose oxidase on the multiple nanowires for determining glucose concentration. The proposed sensor is 340 nm in width and uses five physically identical and parallel nanowires. The sensor contained nanowires of various lengths 3, 5, and 10 μm. Experimental results showed that sensor sensitivity is inversely proportional to nanowire length. The sensor with 3 μm in nanowire length exhibited a theoretical resolution of 0.003 mg-dL and the highest sensitivity of 0.03 μA-mg-dL. Furthermore, the proposed sensor retains this performance when reused for up to 10 applications. View Full-Text

Keywords: multiple nanowires; glucose sensor; enzymatic modification; top-down method multiple nanowires; glucose sensor; enzymatic modification; top-down method





Autor: Cheng-Chih Hsu 1,* , Yu-Ching Liao 1,†, Yen-Ting Tsai 1,†, Hsin-I Yeh 1,† and Chyan-Chyi Wu 2,†

Fuente: http://mdpi.com/



DESCARGAR PDF




Documentos relacionados