Abstract In this study, we report the development of an SPR Surface Plasmon Resonance immunosensor for the detection of ampicillin, operating under flow conditions. SPR sensors based on both direct with the immobilization of the antibody and competitive with the immobilization of the antigen methods did not allow the detection of ampicillin. Therefore, a sandwich-based sensor was developed which showed a good linear response towards ampicillin between 10−3 and 10−1 M, a measurement time of ≤20 min and a high selectivity both towards β-lactam antibiotics and antibiotics of different classes. View Full-Text

Keywords: ampicillin; flow immunosensor SPR; sandwich method ampicillin; flow immunosensor SPR; sandwich method





Autor: Mauro Tomassetti 1,* , Giorgia Conta 1, Luigi Campanella 1, Gabriele Favero 2, Gabriella Sanzò 2, Franco Mazzei 2 and Riccarda Antiochia 2