This article is from Korean Journal of Anesthesiology, volume 65.AbstractMalfunction of the unidirectional valve in a breathing circuit system may cause hypercapnia from the rebreathing of expired gas, ventilation failure, and barotrauma. Capnography is a useful method for monitoring the integrity of the unidirectional valve. We experienced two cases of malfunction of a unidirectional valve which caused leakage and reverse flow, diagnosed early as a change of the capnographic waveform. One case was caused by expiratory unidirectional valve breakage. The other was caused by an incorrectly-assembled inspiratory unidirectional valve.



Autor: Lee, Chol; Lee, Kyu Chang; Kim, Hye Young; Kim, Mi Na; Choi, Eun Kyung; Kim, Ji-Sub; Lee, Won Sang; Lee, Myeong Jong; Kim, Hyung Tae