Methodology, outcome, safety and in vivo accuracy in traditional frame-based stereoelectroencephalographyReportar como inadecuado




Methodology, outcome, safety and in vivo accuracy in traditional frame-based stereoelectroencephalography - Descarga este documento en PDF. Documentación en PDF para descargar gratis. Disponible también para leer online.

Acta Neurochirurgica

pp 1–14

First Online: 05 July 2017Received: 03 April 2017Accepted: 31 May 2017

Abstract

BackgroundStereoelectroencephalography SEEG is an established diagnostic technique for the localization of the epileptogenic zone in drug-resistant epilepsy. In vivo accuracy of SEEG electrode positioning is of paramount importance since higher accuracy may lead to more precise resective surgery, better seizure outcome and reduction of complications.

ObjectiveTo describe experiences with the SEEG technique in our comprehensive epilepsy center, to illustrate surgical methodology, to evaluate in vivo application accuracy and to consider the diagnostic yield of SEEG implantations.

MethodsAll patients who underwent SEEG implantations between September 2008 and April 2016 were analyzed. Planned electrode trajectories were compared with post-implantation trajectories after fusion of pre- and postoperative imaging. Quantitative analysis of deviation using Euclidean distance and directional errors was performed. Explanatory variables for electrode accuracy were analyzed using linear regression modeling. The surgical methodology, procedure-related complications and diagnostic yield were reported.

ResultsSeventy-six implantations were performed in 71 patients, and a total of 902 electrodes were implanted. Median entry and target point deviations were 1.54 mm and 2.93 mm. Several factors that predicted entry and target point accuracy were identified. The rate of major complications was 2.6%. SEEG led to surgical therapy of various modalities in 53 patients 69.7%.

ConclusionsThis study demonstrated that entry and target point localization errors can be predicted by linear regression models, which can aid in identification of high-risk electrode trajectories and further enhancement of accuracy. SEEG is a reliable technique, as demonstrated by the high accuracy of conventional frame-based implantation methodology and the good diagnostic yield.

KeywordsComplications Epilepsy surgery In vivo accuracy Stereoelectroencephalography Stereotaxy  Download fulltext PDF



Autor: Lars E. van der Loo - Olaf E. M. G. Schijns - Govert Hoogland - Albert J. Colon - G. Louis Wagner - Jim T. A. Dings

Fuente: https://link.springer.com/