A Newly Developed Variable Stiffness Duodenoscope for Diagnostic and Therapeutic Endoscopic Retrograde CholangiopancreatographyReportar como inadecuado




A Newly Developed Variable Stiffness Duodenoscope for Diagnostic and Therapeutic Endoscopic Retrograde Cholangiopancreatography - Descarga este documento en PDF. Documentación en PDF para descargar gratis. Disponible también para leer online.

Diagnostic and Therapeutic EndoscopyVolume20102010, Article ID153951, 4 pages

Clinical Study Department of Gastroenterology and Hepatology, Tokyo Medical University, 6-7-1 Nishishinjuku, Shinjuku-ku, Tokyo 160-0023, Japan

Received 3 August 2010; Accepted 28 October 2010

Academic Editor: RamiEliakim

Copyright 2010 Takao Itoi et al.
This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

The aim of this study is to evaluate a prototype variable stiffness duodenoscope VSD for diagnostic and therapeutic ERCP in comparison with standard duodenoscopes.
We performed retrospective analysis on the success rate of intubation of the second duodenum, overall procedural success rate, and comparative frequency of the necessity to change duodenoscopes from standard JF-260V and TJF 260V or to change stiffness using the VSD.
A total of 213 nonconsecutive procedures in 196 patients with pancreaticobiliary diseases.
There was no statistically significant difference in endoscope intubation rate or technical success rate between the different duodenoscopes.
In one patient with severe duodenal stenosis, the VSD using the moderately stiff mode allowed the major papilla to be reached when the TJF-260V endoscope could not.
There were no serious procedure-related adverse events.
In conclusion, while the VSD performed well, the present models do not appear to offer obvious advantages over the standard duodenoscopes for routine diagnostic and therapeutic ERCP.
Prospective studies may be warranted to identify those patients who would benefit from this new technology.





Autor: Takao Itoi,Atsushi Sofuni,Fumihide Itokawa,Toshio Kurihara,Takayoshi Tsuchiya,Kentaro Ishii,Shujiro Tsuji,Nobuhito Ikeuchi,Junko