The objective assessment of cough frequency: accuracy of the LR102 deviceReport as inadecuate




The objective assessment of cough frequency: accuracy of the LR102 device - Download this document for free, or read online. Document in PDF available to download.

Cough

, 7:11

First Online: 01 December 2011Received: 07 February 2011Accepted: 01 December 2011

Abstract

BackgroundThe measurement of cough frequency is problematic and most often based on subjective assessment. The aim of the study was to assess the accuracy of the automatic identification of cough episodes by LR102, a cough frequency meter based on electromyography and audio sensors.

MethodsTen adult patients complaining of cough were recruited in primary care and hospital settings. Participants were asked to wear LR102 for 4 consecutive hours during which they were also filmed.

ResultsMeasures of cough frequency by LR102 and manual counting were closely correlated r = 0.87 for number of cough episodes per hour; r = 0.89 for number of single coughs per hour but LR102 overestimated cough frequency. Bland-Altman plots indicate that differences between the two measurements were not influenced by cough frequency.

ConclusionsLR102 offers a useful estimate of cough frequency in adults in their own environment, while significantly reducing the time required for analysis.

Keywordscough validation study cough meter Electronic supplementary materialThe online version of this article doi:10.1186-1745-9974-7-11 contains supplementary material, which is available to authorized users.